Multi Channel [Home]   [Contact Us]   [Imprint]   [AGB]

MCR-n - Multichannel Spectroradiometer

JETI MCR-n (download datasheet (170kB)) is a multichannel VIS spectroradiometer for the simultaneous measurement of up to 12 light sources. Depending on the application it can be equipped with full or reduced number of channels.
The instrument can be used with the supplied PC software or via serial interface commands. JETI MCR-n can handle up to three calibration data sets for different measuring heads.

    Examples for Applications:
  • Quality control of LED equipped PCB's
  • Goniometric measurements
  • Burn-in tests of lamps

    Advantages:
  • Simultaneously measurement for all channels
  • Continuous measurement mode
  • Measurement in time step intervals
  • Automatic determination of correct integration time
  • Control of the instrument also via interface commands
  • Additional input for hardware triggering of the instrument
  • USB and CAN interface (optional)
  • Data Export to Excel spread sheets
  • Export of calculated measuring values, e.g. chromaticity
    Measuring values:
  • Luminance, Radiance
  • Illuminance, Irradiance
  • xy and u'v' coordinates
  • Dominate wavelength
  • Color purity
  • Correlated Color Temperature (under preparation)
  • Transmission

Specifications
Optical Parameters
Spectral range 380 nm .. 780 nm
Spectral resolution 9 nm (FWHM)
Digital spectral resolution 4 nm
Digital electronic resolution 14 bit A/D
Number of channels 12 (or less)
Accuracy (radiance) ± 10 %
Viewing angle (radiance) approx. 2°
Dispersive Diffraction grating
Light receiving element Photodiode array
Maximum measuring time 60 s
Measuring values Spectral radiance;
Integral luminance radiance;
Chromaticity coordinates x,y; u',v';
Correlated Color Temperature (under preparation);
Dominant wavelength, color purity
Transmission (with additional light source)
Measuring range and accuracy
Luminance 2 .. 7×104 cd/m2
Reproducibility (radiance) ± 1 % (new measurement)
± 5 % (after new start)
Luminance accuracy ± 5 %
Luminance reproducibility ± 2 %
Chromaticity accuracy ± 0.002 x, y (@ 2856 K)
Color reproducibility ± 0.0005 x, y
CCT reproducibility ± 20 K (@ 2856 K)
Technical parameters
Operating conditions Temperature 10 .. 40 °C
Humidity < 85 % relative humidity at 35 °C
PC interfaces RS 232, 8N1, 115200 Bd or optional
USB1.1 and USB2.0 compatible up to 1 Mbyte/sec
CAN optional
Dimensions 300 mm x 250 mm x 90 mm
Weight approx. 1000 g
Power supply 230 VAC (115 VAC optional)/ 9 VDC (power supply included)
Accessories (included) PC software JETI MCR-n for Windows 2000/XP /NT
Operation manual
Measuring heads Application specific

© 1998 .. 2010 JETI Technische Instrumente GmbH -- Tatzendpromenade 2 -- D-07745 Jena -- Germany -- March 2010