The size of the integrating sphere depends from the measuring object. A baffle avoids the inclusion of direct radiation from the light source to the measurement.
Nominal sphere diameters are 50, 100, 150, 250 and 500 mm. Other sphere sizes are possible.
Specification
| Optical Parameters |
| Spectral range |
380 nm .. 780 nm |
| Optical bandwith |
5 nm |
| Wavelength resolution |
1 nm |
| Digital electronic resolution |
16 bit ADC (15 bit used) |
| Dispersive element |
Diffraction grating (9nm FWHM) |
| Light receiving element |
Photodiode array 1024 pixel (binned to 256 pixel) |
| Measuring values |
Spectral Radiant Flux;
Total Radiant Flux/ Luminous Flux
Chromaticity coordinates x,y; u',v';
Correlated Color Temperature;
Dominant wavelength, color purity;
Color Rendering Index
|
| Measuring range and accuracies |
| Measuring range Luminous Flux |
1 lm ... 4000 lm (depending from integrating sphere, see Application Note 5) |
| Luminous Flux accuracy |
depending from integrating sphere |
| Luminous Flux reproducibility |
depending from integrating sphere |
| Chromaticity accuracy |
± 0.002 x, y (@ 2856 K) |
| Color reproducibility |
± 0.0005 x, y |
| CCT reproducibility |
± 20 K (@ 2856 K) |
| Wavelength accuracy |
± 0.5mm |
| Technical Parameters |
| Sphere Diameter |
50-150 mm |
| PC interface |
USB2.0 fullspeed |
| Operating conditions |
Temperature 10 ... 40 °C
Humidity < 85 % relative humidity at 35 °C
|
| Dimensions |
125 mm x 58 mm x 34 mm |
| Weight |
550 g |
| Power Supply |
Hub powered |
| Miscellaneous |
| Accessories (included) |
PC software JETI LiMeS for Windows 2000/XP/Vista
Integration Sphere
Operation manual
USB cable
Calibration certificate
|
| NIST traceable calibration |
Recommended recalibration interval: one year |