specbos 1401 - Spectroradiometer
|
specbos 1401 is a compact luminous intensity meter, especially for the measurement of the averaged intensity according to the CIE127 publication (download datasheet (103kB)).
It is delivered with a measuring head for CIE cond. B and an extension tube for cond. A.
Its easy-to-use software has the full complement of radiometric and colorimetric functions requisite for research, diagnostic, and quality control applications.
|
|
|
Measuring values:
- Radiant Intensity, Luminous Intensity
- Spectral Radiant Intensity
- xy and u'v' coordinates
- Dominant wavelength
- Color purity
- Correlated Color Temperature
- Color Rendering Index
|
Sort your LEDs with the LED Ranker.
For further information see LED Ranker site with demo video.
Advantages:
- Easy change between condition A and B
- USB powered - no extra power supply
- Automatic determination of measuring time
- Shutter for dark correction
- Data export into Excel spread sheets
|
Applications:
- LED characterization
- Luminous Intensity measurement of lamps and luminaires
|
Specification
| Optical Parameters |
| Spectral range |
380 nm .. 780 nm |
| Optical bandwith |
5 nm |
| Wavelength resolution |
1 nm |
| Digital electronic resolution |
16 bit ADC (15 Bit used) |
| Dispersive Element |
Diffraction grating |
| Light Receiving Element |
Photodiode array 1024 pixel (binned to 256 pixel) |
| Measuring values |
Spectral Radiant Intensity accord. to CIE127 Cond. A/B;
Total Radiant Intensity/ Luminous Intensity;
Chromaticity coordinates x,y; u',v';
Correlated Color Temperature;
Dominant Wavelength, color purity;
Color Rendering Index;
|
| Measuring range and accuracies |
| Measuring range Lum. Intensity |
50 mcd .. 50 cd |
| Luminous intensity accuracy |
t.b.d |
| Luminous Intensity reproducibility |
t.b.d |
| Chromaticity accuracy |
± 0.002 x, y (@2856K) |
| Color reproducibility |
± 0.0005 x, y |
| CCT reproducibility |
± 20 K (@ 2856 K) |
| Wavelength accuracy |
± 0.5 nm |
| Technical Parameters |
| Interface |
USB 2.0 fullspeed |
| Operating Conditions |
| Temperature |
10 .. 40 °C |
| Humidity |
< 85% relative humidity @ 35 °C |
|
| Power supply |
USB powered |
| Miscellaneous |
| Accessories (included) |
Measuring tubes for Cond. A and B
Cosine diffusor (for irradiance measurement)
PC software JETI LiMeS for Windows 2000/XP/Vista
DLL, LabView VI's
Operation manual
Calibration certificate
USB cable
|
| NIST traceable Calibration |
Recommended interval: 1 year |
|