Specification
| Optical Parameters |
| Spectral range |
400 nm .. 780 nm |
| Measuring range and accuracies |
| Wavelength accuracy |
± 0.5 nm (ASTM E275, filter BG 20, 2 mm, at 528.7 nm/ 684.3 nm) |
| Wavelength reproducibility |
± 0.2 nm (ASTM E275, filter BG 20, 2 mm, at 528.7nm/ 684.3 nm) |
| Photometric range |
0 .. 3 AU (ASTM E275, ND filters, at 550 nm) |
| Photometric precision |
± 0.005 AU (ASTM E275, D = 0.46, at 550 nm) |
| Photometric accuracy |
± 0.002 AU (ASTM E275, D = 1, at 550 nm) |
| Stray light |
< 10-3 (ASTM E387, GG 495, 2 mm, at 430 nm) |
| Sensitivity |
typ. 1.3×1014 counts/Ws (550 nm) |
| Technical Parameters |
| Input |
SMA (optional ST) |
| Slit |
100 µm fiber (optional 50 µm), NA 0.22 |
| Light source |
Lamp + blue LED
|
| Integration time |
10 ... 60 000 ms |
| Operating cond. |
Temperature 10 ... 40 °C
Humidity < 85 % relative humidity at 35 °C
|
| Miscellaneous |
| Applications |
Measurement of reflection and transmission spectra, e.g. of solid surfaces, filters and liquids
|
Accessories (included)
|
PC software JETI VersaSpec for Windows 2000/XP/ Vista (since VersaSpec Release 3.1.5)
Operation manual
USB cable
Transport box
|
| Dispersive element |
Diffraction grating
|
| Light receiving element |
128 pixel photodiode array |