Spectrophotoometer [Home]   [Contact Us]   [Imprint]

Spectrophotometer

specbos 3001

specbos 3001 (download datasheet (172kB)) is an easy to use instrument for spectral transmission measurement in the VIS region. It is especially suited for test cells of up to 10 mm optical path length and for flat materials like optical filters.
The software package allows the measurement of spectral data as transmission and absorbance. Besides the display and management of spectra (hot links to Excel and ASCII files) further mathematical operations as first and second derivation are possible.
The optional software JETI specbos color allows the measurement of color coordinates of samples (e.g. L*a*b*).

Absorption spectrum of an optical filter (BG 36) click to enlarge

Convince yourself by this new type of photometer! Easy to handle, reliable and compact.

Specification
Samples Filters, test cells (1 .. 10 mm optical path)
Measuring geometry 0/0 (illumination by 0°, observation by 180°)
Centre height 15 mm
Measuring diameter 3 mm
Measuring time approx. 0.7 s
Spectral range 400 .. 780 nm
Light source Lamp/ LED
Dispersive element Diffraction grating
Light receiving element 128 pixel photodiode array
Operating cond. Temperature 10 ... 40 °C
Humidity < 85 % relative humidity at 35 °C
Power supply Hub powered
PC interfaces USB2.0 fullspeed (12Mbit/s)
Applications Measurement of reflection and transmission spectra, e.g. of mirrors
Accessories (included) PC software VersaSpec for Windows 2000/XP/ Vista (since VersaSpec Release 3.1.5)
Operation manual
Transport box

Photometric measurement with Software: VersaSpec
Measurement curves Absorbance spectrum
Wavelength accuracy ± 0.5 nm (ASTM E275, filter BG 20, 2 mm, at 528.7 nm/ 684.3 nm)
Wavelength reproducibility ± 0.2 nm (ASTM E275, filter BG 20, 2 mm, at 528.7 nm/ 684.3 nm)
Base line noise max. 10-3 AU at 550 nm (water filled 1 cm test cell)
Base line drift max. 2×10-3
Photometric precision ± 0.002 AU (ASTM E275, D = 1, at 550 nm)
Photometric accuracy ± 0.005 AU (ASTM E275, D = 0.46, at 550 nm)
Photometric linearity 0 ... 2.6 AU (ASTM E275, ND filters, at 550 nm)
Stray light < 0.1% (ASTM E387, GG495, 4 mm, at 420 nm/ 630 nm)
Long term stability ± 0.3 % / h
Temperature dependence < 5×10-4 AU/ K (D = 0.2, at 550 nm)
Option Kinetics mode (under preparation)

Color measurement with optional Software: JETI specbos color
Wavelength pitch 10 nm
Illumination C (A and D65 selectable)
Observer 2 ° (10 ° selectable)
Measuring values Color systems X, Y, Z; L*a*b*(CIE), Luv, L*C*h*
Color differences
Jod, Hazen, Gardner (under preparation)
Precision dE < 0.1