Reflexion measurement of displays

Recommended products: specbos 2501 with add-on optics and special integrating sphere

specbos 2501 can be used for reflexion measurements, if an illumination source and a set up realizing the appropriate measuring geometry will be added. In addition, the accessories add-on optics and integrating sphere are required.

An example is the measurement of display reflexion, as described e.g. in the Information Display Measurements standard¹.

An integrating sphere with lamp(s) illuminating its interior is used to create a homogenious and diffuse illumination of the display, which is mounted in the centre of the sphere. The measurement can be done with specular reflex uncluded or excluded. The angle θd is normally 8°.

display reflexion mitspiegelreflex

Display reflexion measurement with specular reflex included

display reflexion ohnespiegelreflex

Display reflexion measurement with specular reflex excluded

 

The following measurements are performed with switched on display:
  • Measurement of the display Luminance when illuminated by the sphere illumination (LQon)
  • Measurement of the reflectance standard Luminance when illuminated by the sphere illumination (LstdQon)
  • Measurement of the reflectance standard Luminance without sphere illumination (LstdQQoff)
  • Measurement of the display Luminance when sphere illumination is off (LQoff)

The resulting reflectance ρ8/di (specular reflex included) and ρ8/di (specular reflex included) will be calculated as follows: